Tiger-i 2000 HCl

Monitor for Ambient Molecular Contaminants (AMCs)

Applications

  • Air Quality
  • Airborne Molecular Contaminants
  • Research & Development
Detection and Matrix Range LDL (3σ) LDL (peak-to-peak)
HCl in Ambient Air0 – 4 ppm0.75 ppb1.0 ppb
HCl in Clean Dry Air (CDA)0 – 4 ppm0.75 ppb1.0 ppb
HCl in N20 – 4 ppm0.75 ppb1.0 ppb
Performance
Operation range See Detection Capability table
Detection limit (LDL, 3σ or 24 h peak-to-peak) See Detection Capability table
Precision (1σ, greater of) ± 0.75% or 1/3 of LDL
Accuracy (greater of) ± 4% or the LDL
Zero drift None
Speed of response 2 minutes to 80%
Environmental conditions 10°C to 40°C, 30% to 80% RH (non-condensing)
Storage temperature -10° to 50°C
Gas Handling System and Conditions*
Wetted materials Sample gas compatible
Gas connections 1/4" male VCR inlet & outlet
Leak tested to 1 x 10-9 mbar l /sec
Inlet pressure 0 – 15 psig (1 – 2 bara)
Outlet pressure Vacuum (<10 Torr)
Flow rate ∼1 slpm
Sample gases Ambient air, clean dry air (CDA), or N2
Gas temperature Up to 60°C
Dimensions
Standard sensor 8.73" x 8.57" x 23.6" (222 mm x 218 mm x 599 mm)
Sensor rack 8.73" x 19.0" x 23.6" (222 mm x 483 mm x 599 mm)
Weight
Standard sensor 33 lbs (15 kg)
AMC cart (excl. sensors and pumps) 260 lbs (118 kg)
Electrical
Alarm indicators 2 user programmable, 1 system fault, Form C relays
Power requirements 90 – 240 VAC 50/60 Hz
Power consumption 40 Watts max. per sensor
Signal output Isolated 4–20 mA per sensor
User interfaces 5.7" LCD touchscreen, 10/100 Base-T Ethernet, RS-232
  802.11g Wireless (optional), Modbus TCP (optional)
Patents
U.S. Patent #7,277,177  
   
*Vacuum source with >2 slpm @ 10 Torr required  