Specification Tables

Summary of Analyzer Specifications and Detection Capabilities

Please select a product family.

Tiger-i Specification Summary

Product Detection and Matrix Range LDL (3σ) LDL (peak-to-peak) Accuracy (greater of) Precision (greater of) Flow rate Pressure
Tiger-i 2000 HF
Monitor for Airborne Molecular Contaminants (AMCs)
HF in Ambient Air 0 – 1 ppm 0.15 ppb 0.2 ppb ± 4% or the LDL ± 0.75% or ⅓ LDL ∼1 slpm 0 – 15 psig
HF in Clean Dry Air (CDA) 0 – 1 ppm 0.15 ppb 0.2 ppb
HF in N2 0 – 1 ppm 0.15 ppb 0.2 ppb
Product Detection and Matrix Range LDL (3σ) LDL (peak-to-peak) Accuracy (greater of) Precision (greater of) Flow rate Pressure
Tiger-i 2000 HCl
Monitor for Ambient Molecular Contaminants (AMCs)
HCl in Ambient Air 0 – 4 ppm 0.75 ppb 1.0 ppb ± 4% or the LDL ± 0.75% or ⅓ LDL ∼1 slpm 0 – 15 psig
HCl in Clean Dry Air (CDA) 0 – 4 ppm 0.75 ppb 1.0 ppb
HCl in N2 0 – 4 ppm 0.75 ppb 1.0 ppb
Product Detection and Matrix Range LDL (3σ) LDL (peak-to-peak) Accuracy (greater of) Precision (greater of) Flow rate Pressure
Tiger-i 2000 NH3
Monitor for Ambient Molecular Contaminants (AMCs)
NH3 in Ambient Air 0 – 40 ppm 2.5 ppb 3.5 ppb ± 4% or the LDL ± 0.75% or ⅓ LDL ∼1 slpm 0 – 15 psig
NH3 in Clean Dry Air (CDA) 0 – 40 ppm 2.5 ppb 3.5 ppb
NH3 in N2 0 – 40 ppm 2.5 ppb 3.5 ppb
LDL achievable with humidity
variation of less than 7% RH
Product Detection and Matrix Range LDL Sensitivity Accuracy (greater of) Precision (greater of) Flow rate Pressure
Tiger-i 2000 CH4
Methane analyzer for greenhouse gas detection and monitoring
CH4 in Clean Dry Air (CDA) (low range) 0 – 20 ppm 2.0 ppb 1.5 ppb ± 4% or the LDL
(∼80 ppb at ambient CH4 levels)
± 0.75% or ⅓ Sensitivity
(∼15 ppb at ambient CH4 levels)
∼1 slpm 0 – 15 psig
CH4 in Clean Dry Air (CDA) (high range) 0 – 100 ppm 10 ppb 7.5 ppb
CH4 in N2 (low range) 0 – 20 ppm 2.0 ppb 1.5 ppb
CH4 in N2 (high range) 0 – 100 ppm 10 ppb 7.5 ppb
CH4 in Ambient Air (low range) 0 – 25 ppm See Resources Brochure See Resources Brochure
CH4 in Ambient Air (high range) 0 – 100 ppm See Resources Brochure See Resources Brochure
Upper range as high as 1000 ppm available on request
Product Detection and Matrix Range LDL Sensitivity Accuracy (greater of) Precision (greater of) Flow rate Pressure
Tiger-i 2000 CO2
CO2 analyzer for greenhouse gas monitoring
CO2 in Clean Dry Air (CDA) 0 – 3000 ppm 200 ppb 150 ppb ± 4% or the LDL
(∼16 ppm at ambient CO2 levels)
± 0.75% or ⅓ Sensitivity
(∼3 ppm at ambient CO2 levels)
∼1 slpm 0 – 15 psig
CO2 in N2 0 – 3000 ppm 200 ppb 150 ppb
CO2 in Ambient Air 0 – 1500 ppm N/A N/A
Product Detection and Matrix Range LDL Sensitivity Accuracy (greater of) Precision (greater of) Flow rate Pressure
Tiger-i 2000 CH2O
Trace formaldehyde in ambient air
CH2O in Ambient Air 0 – 70 ppm 7 ppb 5 ppb ± 4% or the LDL ± 0.75% or 1/3 Sensitivity ∼1 slpm 0 – 15 psig
Product Model Range LDL (peak-to-peak) Sensitivity (3σ) Accuracy (greater of) Precision (greater of) Flow rate Pressure
Tiger-i CEM
Continuous Emissions Monitoring (CEM) for HCl, HF, NH3 and more
Tiger-i CEM NH3 0 – 40 ppm 8 ppb 6 ppb ± 4% or the LDL ± 0.75% or 1/3 Sensitivity ∼1 slpm 0 – 10 psig (HCl and HF)
0 – 15 psig (all others)
Tiger-i CEM HCl 0 – 4 ppm 1.0 ppb 0.75 ppb
Tiger-i CEM HF 0 – 1 ppm 0.2 ppb 0.15 ppb
Tiger-i CEM H2S 0 – 500 ppm 50 ppb 38 ppb
Tiger-i CEM CO 0 – 2000 ppm 100 ppb 75 ppb
Tiger-i CEM HCN 0 – 15 ppm 1.5 ppb 1.1 ppb
Tiger-i CEM CH4 0 – 20 ppm 2.0 ppb 1.5 ppb