SPARK C2H2

trace-level acetylene detection

SPECIFICATIONS

Performance
Operating range See Detection Capability table
Detection Limit (3σ/24h) See Detection Capability table
Precision (1σ, greater of) ± 0.75% or 1/3 of LDL
Accuracy (greater of) ± 4% or LDL
Speed of response < 1 minute to 90%
Environmental conditions 10°C to 40°C, 30% to 80% RH (non-condensing)
Storage temperature -10°C to 50°C
Gas Handling System and Conditions
Wetted materials 316L stainless steel, 10 Ra surface finish
Gas connections 1/4″ male VCR inlet and outlet
Inlet pressure 10 – 125 psig (1.7 – 9.6 bar)
Flow rate ∼0.7 slpm (in N2), gas-dependent
Sample gases Most inert, toxic, and passive matrices
Gas temperature Up to 60°C
Dimensions, H x W x D
Standard sensor 8.73″ x 8.57″ x 23.6″ (222 mm x 218 mm x 599 mm)
Sensor rack (fits up to two sensors) 8.73″ x 19.0″ x 23.6″ (222 mm x 483 mm x 599 mm)
Weight
Standard sensor 32 lbs (14.5 kg)
Electrical and Interfaces
Platform Max series analyzer
Alarm indicators 2 user programmable, 1 system fault, Form C relays
Power requirements 90 – 240 VAC, 50/60 Hz
Power consumption 40 Watts max.
Signal output Isolated 4–20 mA per sensor
User interfaces 5.7″ LCD touch screen, 10/100 Base-T Ethernet
  USB, RS-232, RS-485, Modbus TCP (optional)
Data storage Internal and external flash drive
Certification CE Mark
Patents
U.S. Patent #7,277,177  