T-I Max AIR CH4

analyzer for greenhouse gas monitoring

  • Greenhouse Gas Monitoring
  • Research & Development

DETECTION CAPABILITY

Detection and Matrix                  Range          LDL (3σ)       Precision (1σ) @ zero
Low-Range Model      
CH4 in Clean Dry Air (CDA) 0 – 25 ppm 1.5 ppb 0.5 ppb
CH4 in N2 0 – 25 ppm 1.5 ppb 0.5 ppb
CH4 in Ambient Air 0 – 25 ppm N/A N/A
High-Range Model      
CH4 in Clean Dry Air (CDA) 0 – 100 ppm 7.5 ppb 2.5 ppb
CH4 in N2 0 – 100 ppm 7.5 ppb 2.5 ppb
CH4 in Ambient Air 0 – 100 ppm N/A N/A
Upper range available as high as
1000 ppm on request
     

SPECIFICATIONS

Performance
Operating range See Detection Capability table
Detection limit (LDL, 3σ/24h) See Detection Capability table
Precision (1σ, greater of) ± 0.75% or 1/3 of LDL
  (equivalent to 15 ppb over 24 h at typical ambient CH4 levels)
Accuracy (greater of) ± 4% or the LDL
  (equivalent to 80 ppb over 24 h at typical ambient CH4 levels)
Speed of response 3 minutes to 95%
Environmental conditions 10°C to 40°C, 30% to 80% RH (non-condensing)
Storage temperature -10°C to 50°C
Gas Handling System and Conditions*
Wetted materials 316L stainless steel, 10 Ra surface finish
Gas connections 1/4″ male VCR inlet & outlet
Leak tested to 1 x 10-9 mbar l / sec
Inlet pressure 0 – 15 psig (1 – 2 bara)
Outlet pressure Vacuum (<10 Torr)
Flow rate ∼1 slpm
Sample gases Ambient air, clean dry air (CDA) and N2
Gas temperature Up to 60°C
Dimensions, H x W x D
Standard sensors 8.73″ x 8.57″ x 23.6″ (222 mm x 218 mm x 599 mm)
Sensor rack (fits up to two sensors) 8.73″ x 19.0″ x 23.6″ (222 mm x 483 mm x 599 mm)
Weight
Standard sensor 33 lbs (15 kg)
Electrical and Interfaces
Platform Max series analyzer
Alarm indicators 2 user programmable, 1 system fault, Form-C relay
Power requirements 90 − 240 VAC, 50/60 Hz
Power consumption 40 Watts max.
Signal output Isolated 4–20 mA per sensor
User interfaces 5.7″ LCD touchscreen, 10/100 Base-T Ethernet
  USB, RS-232, RS-485, Modbus TCP (optional)
Data storage Internal or external flash drive
Certification CE Mark
Patents
U.S. Patent #7,277,177  
   

*Vacuum source with >2 slpm @ 10 Torr required